A 7 GHz Biosensor for permittivity change with enhanced sensitivity through phase compensation
J. Wessel, K. Schmalz, C. Meliani, B. Cahill, J.C. Scheytt, in: Institute of Electrical and Electronics Engineers (IEEE) (Ed.), 2014 44th European Microwave Conference, IEEE, 2014, pp. 699–702.
Download
Es wurde kein Volltext hochgeladen. Nur Publikationsnachweis!
Konferenzbeitrag
| Veröffentlicht
| Englisch
Autor*in
Wessel, Jan;
Schmalz, Klaus;
Meliani, Chafik;
Cahill, Brian
;
Scheytt, J. Christoph
herausgebende Körperschaft
Institute of Electrical and Electronics Engineers (IEEE)
Abstract
A calibration technique as well as measurement results for a 7 GHz Biosensor are presented. It is shown that the applied sensor structure can be calibrated by adjusting the phase of a sensing element's transmission S21. This is realized by slowing down the wave traveling a microstrip line serving as a reference in the differential sensor structure. The dielectric properties along with certain physical boundaries of an obstacle covering parts of the microstrip line evoke that effect. Measurements with an ethanol serious along with simulation results showed that sensitivity can be increased substantially with this calibration technique. A change of the real part of the sample's permittivity of 48 leads to a 18 MHz frequency shift.
Erscheinungsjahr
Titel des Konferenzbandes
2014 44th European Microwave Conference
Seite
699-702
Konferenz
2014 44th European Microwave Conference (EuMC)
Konferenzort
Rome
Konferenzdatum
2014-10-06 – 2014-10-09
FH-PUB-ID
Zitieren
Wessel, Jan ; Schmalz, Klaus ; Meliani, Chafik ; Cahill, Brian ; Scheytt, J. Christoph: A 7 GHz Biosensor for permittivity change with enhanced sensitivity through phase compensation. In: Institute of Electrical and Electronics Engineers (IEEE) (Hrsg.): 2014 44th European Microwave Conference : IEEE, 2014, S. 699–702
Wessel J, Schmalz K, Meliani C, Cahill B, Scheytt JC. A 7 GHz Biosensor for permittivity change with enhanced sensitivity through phase compensation. In: Institute of Electrical and Electronics Engineers (IEEE), ed. 2014 44th European Microwave Conference. IEEE; 2014:699-702. doi:10.1109/EuMC.2014.6986530
Wessel, J., Schmalz, K., Meliani, C., Cahill, B., & Scheytt, J. C. (2014). A 7 GHz Biosensor for permittivity change with enhanced sensitivity through phase compensation. In Institute of Electrical and Electronics Engineers (IEEE) (Ed.), 2014 44th European Microwave Conference (pp. 699–702). Rome : IEEE. https://doi.org/10.1109/EuMC.2014.6986530
@inproceedings{Wessel_Schmalz_Meliani_Cahill_Scheytt_2014, title={A 7 GHz Biosensor for permittivity change with enhanced sensitivity through phase compensation}, DOI={10.1109/EuMC.2014.6986530}, booktitle={2014 44th European Microwave Conference}, publisher={IEEE}, author={Wessel, Jan and Schmalz, Klaus and Meliani, Chafik and Cahill, Brian and Scheytt, J. Christoph}, editor={Institute of Electrical and Electronics Engineers (IEEE)Editor}, year={2014}, pages={699–702} }
Wessel, Jan, Klaus Schmalz, Chafik Meliani, Brian Cahill, and J. Christoph Scheytt. “A 7 GHz Biosensor for Permittivity Change with Enhanced Sensitivity through Phase Compensation.” In 2014 44th European Microwave Conference, edited by Institute of Electrical and Electronics Engineers (IEEE), 699–702. IEEE, 2014. https://doi.org/10.1109/EuMC.2014.6986530.
J. Wessel, K. Schmalz, C. Meliani, B. Cahill, and J. C. Scheytt, “A 7 GHz Biosensor for permittivity change with enhanced sensitivity through phase compensation,” in 2014 44th European Microwave Conference, Rome , 2014, pp. 699–702.
Wessel, Jan, et al. “A 7 GHz Biosensor for Permittivity Change with Enhanced Sensitivity through Phase Compensation.” 2014 44th European Microwave Conference, edited by Institute of Electrical and Electronics Engineers (IEEE), IEEE, 2014, pp. 699–702, doi:10.1109/EuMC.2014.6986530.